Testing with Class 1 Sound Calibrator

Estimated reading: 2 minutes 559 views

Summary: This page documents the results of testing the PCBA-DBM-r3 module with a Class 1 sound level calibrator.

This article demonstrates the accuracy of the PCBA-DBM-r3 by testing it with a sound level calibrator.
The calibrator produces 94 dB SPL test tones at 1 kHz, 500 Hz, 250 Hz, and 125 Hz to test the module’s measurement and weighting accuracy at different frequencies.
The acoustic calibrator used (AWA6223+ multi-frequency calibrator) is IEC 60942:2017 Class 1 compliant and dynamically compensates for environmental factors like temperature, humidity and atmospheric pressure for accurate calibration tone generation.

💡 TIP
This article on Frequency Weightings by NTI Audio lists “frequency vs weights” used for A-, C-, and Z-weighting in accordance with the IEC 61672:2014 standard.

Test Setup

For testing, the PCBA-DBM-r3 sound level meter module is attached to the AWA6223+ acoustic calibrator using an adapter and sealing gasket. The readings produced by the PCBA-DBM-r3 are read and displayed on demo hardware with a display. The test setup is shown in the picture below.

PCBA-DBM-r3 Acoustic Calibrator Measurement Adapter
Measurement Adapter Setup for Sound Level Calibrator

The touch GUI has buttons to select frequency weights for easy testing.

A-Weighting Mode

Test Tone (94 dB SPL) Ideal Reading (IEC-61672) PCBA-DBM-r3 Reading
1000 Hz 94.0 94
500 Hz 90.8 91
250 Hz 85.4 85
125 Hz 77.9 78
PCBA-DBM-r3 A-weighted 1000 Hz sound level
A-Weighted Reading at 1 kHz
PCBA-DBM-r3 A-weighted 500 Hz sound level
A-Weighted Reading at 500 Hz
PCBA-DBM-r3 A-weighted 250 Hz sound level
A-Weighted Reading at 250 Hz
PCBA-DBM-r3 A-weighted 125 Hz sound level
A-Weighted Reading at 125 Hz

Similarly, readings for C-weighting and Z-weighting were also recorded and are listed below.

C-Weighting Mode

Test Tone (94 dB SPL) Ideal Reading (IEC-61672) PCBA-DBM-r3 Reading
1000 Hz 94.0 94
500 Hz 94.0 94
250 Hz 94.0 94
125 Hz 93.8 94

Z-Weighting Mode

Test Tone (94 dB SPL) Ideal Reading (IEC-61672) PCBA-DBM-r3 Reading
1000 Hz 94.0 94
500 Hz 94.0 94
250 Hz 94.0 94
125 Hz 94.0 94

Revision History

September 12, 2025
– Initial release


Share this Doc